Al Atawneh S. Analytical Formula for the total scatter factor of the Small Radiation Field Size. Journal of Research in Applied and Basic Medical Sciences 2021; 7 (3) :152-160
URL: http://ijrabms.umsu.ac.ir/article-1-168-en.html
Department of Radiation Oncology, King Hussein Cancer Center, Amman, Jordan , saed.al-atawneh@atomki.hu
Abstract:   (1136 Views)
Background & Aims:  We present an analytical formula (AF) to calculate the total scatter factor of the MLC-shaped small radiation beams.
Materials & Methods: Pinpoint, Semiflex, and Farmer ion chambers were used for MLC/Jaw-shaped fields extending from (10x10 cm2) down to (1x1 cm2). The total scatters factor, beam profiles, and penumbra was investigated for MLC/Jaw-shaped beams.
Results: We found that the total scatters factor and the penumbra was clearly higher for MLC-shaped beams due to the MLC leakage, which becomes more and more significant for the small fields. The Analytical Formula was introduced as a least-square fit equation and verified against to measure data. The maximum deviation at small field size (< 4 cm2) was within 3.6% and 43.9% from both MLC and Jaw-shaped beams, respectively.
Conclusion: Our analytical formula shows a very good agreement with measured data at a large field size (> 4 cm2), and had small deviation with MLC-shaped small field size, emphasizing that our analytical formula is valid and reasonably good for the total scatters factor calculation of the MLC-shaped beams, especially the ultra-small field size.
Full-Text [PDF 813 kb]   (740 Downloads)    
Type of Study: orginal article | Subject: Special

References
1. Masanga W, Tangboonduangjit P, Khamfongkhruea C, Tannanonta C. Determination of small field output factors in 6 and 10 MV flattening filter free photon beams using various detectors. J. Phys. Conf. Ser 2016;694:012027. [DOI:10.1088/1742-6596/694/1/012027]
2. Das IJ, Ding GX, Ahnesjo A. Small fields: nonequilibrium radiation dosimetry. Medical physics 2008;35(1):206-15. [DOI:10.1118/1.2815356] [PMID]
3. Charles PH, Cranmer-Sargison G, Thwaites DI, Crowe SB, Kairn T, Knight RT, et al. A practical and theoretical definition of very small field size for radiotherapy output factor measurements. Medical physics 2014;41(4):041707. [DOI:10.1118/1.4868461] [PMID]
4. Klein DM, Tailor RC, Archambault L, Wang L, Therriault-Proulx F, Beddar AS. Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors. Medical physics 2010;37(10):5541-9. [DOI:10.1118/1.3488981] [PMID] [PMCID]
5. Zhu TC, Ahnesjö A, Lam KL, Li XA, Ma C-MC, Palta JR, et al. Report of AAPM Therapy Physics Committee Task Group 74: In-air output ratio, , for megavoltage photon beams. Medical physics 2009;36(11):5261-91. [DOI:10.1118/1.3227367] [PMID]
6. Huq MS, Hwang M-S, Teo TP, Jang SY, Heron DE, Lalonde RJ. A dosimetric evaluation of the IAEA-AAPM TRS483 code of practice for dosimetry of small static fields used in conventional linac beams and comparison with IAEA TRS-398, AAPM TG51, and TG51 Addendum protocols. Medical physics 2018;45(9):4257-73. [DOI:10.1002/mp.13092] [PMID]
7. Laub WU, Wong T. The volume effect of detectors in the dosimetry of small fields used in IMRT. Medical physics 2003;30(3):341-7. [DOI:10.1118/1.1544678] [PMID]
8. Francescon P, Cora S, Cavedon C. Total scatter factors of small beams: a multidetector and Monte Carlo study. Medical physics 2008;35(2):504-13. [DOI:10.1118/1.2828195] [PMID] [PMCID]
9. Klein DM, Tailor RC, Archambault L, Wang L, Therriault-Proulx F, Beddar AS. Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors. Medical physics 2010;37(10):5541-9. [DOI:10.1118/1.3488981] [PMID] [PMCID]
10. Palmans H, Andreo P, Huq MS, Seuntjens J, Christaki KE, Meghzifene A. Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA-AAPM international Code of Practice for reference and relative dose determination. Medical physics 2018;45(11):e1123-e45. [DOI:10.1002/mp.13208]
11. Sauer OA, Wilbert J. Measurement of output factors for small photon beams. Medical physics. 2007;34(6):1983-8. [DOI:10.1118/1.2734383] [PMID]
12. Hrbacek J, Lang S, Klock S. Commissioning of photon beams of a flattening filter-free linear accelerator and the accuracy of beam modeling using an anisotropic analytical algorithm. Int. J. Radiat. Oncol. Biol. Phys 2011;80(4):1228-37. [DOI:10.1016/j.ijrobp.2010.09.050] [PMID]

Add your comments about this article : Your username or Email:
CAPTCHA

Send email to the article author


Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.

© 2024 CC BY-NC 4.0 | Journal of Research in Applied and Basic Medical Sciences

Designed & Developed by : Yektaweb